ATI’s maximum photometer recalibration interval is one year. While ATI can issue calibration certification for less than 12 months upon request, we cannot issue certification for intervals longer than 12 months.
This one year interval may be extended at the discretion of the customer with documentation and control by their respective internal QA department.