Material | Block Temperature | Internal Pressure | Flow | Photometric Concentration | Results/Comments |
---|---|---|---|---|---|
DOP/DEHP | 750-760F | 5 psig | 10-11 lpm | 60% with metering valve fully open | Aerosol stream stable, uniform and dry. Output remained stable and dry while varying the concentration across operating range |
DOS/DEHS | 780-790F | 4.75 psig | 6 lpm | 75% | Aerosol stream dry and uniform. Clogging of the heater block occurred due to increased block temperature |
PAO-4 (Emery 3004) | 760-775F | 5 psig | 7 lpm | 75% | Aerosol stream dry, uniform and stable throughout entire operating range. Increases in inert gas flow resulted in wet aerosol. Approved by FDA, DOE, US Army, Surgeon General as DOP replacement |
Ondina Oil | 760-780F | 5 psig | 7 lpm | 75% | Aerosol stream is dry and stable. Wet aerosol occurred at < 670 F. Aerosol pulsation occurred at > 900°F. GOOD REPLACEMENT FOR DOP. |
Corn Oil | 750-760F | 5 psig | 10 lpm | N/A | Aerosol stream wet. Increasing temperature and reducing inert gas flow resulted in no aerosol output at all. Not Acceptable. |
Mineral Oil | 750-760F | 5 psig | 10 lpm | N/A | Aerosol stream semi-wet. Increasing the temperature while reducing inert gas flow dried aerosol but caused rapid clogging of heater block |