Introduction:
The 100Xp Automated Filter Tester is designed to meet the oil aerosol testing requirement of NIOSH 42 CFR Part 84. This testing specification requires an oil aerosol (DOP or equivalent) with a count median diameter (CMD) of 0.185 ± 0.020 µm with a particle distribution having a standard geometric deviation of ≤1.6. The aerosol produced is also subjected to an ionized air stream to shift the electrically charged generated aerosol to a neutral state (Boltzmann Equilibrium) characteristic of naturally occurring aerosols.
Schematic Diagram of the 100Xp Automated Filter Tester
Aerosol Generation:
Aerosol generation in the 100Xp is achieved using an ingenious, patented submerged nozzle nebulizer and regeneration technique. An air jet exiting the nozzle at high speed below the surface of the aerosol liquid generates bubbles that remain suspended within the liquid. When these bubbles reach the surface of the liquid they burst and generate airborne oil particles necessary to perform efficiency testing. Aerosol generation occurs at a level exceeding the maximum test flow capabilities of the 100Xp. The purpose of this excess aerosol is to maintain constant levels of aerosol regardless of the sample rate. Excess aerosol is exhausted from the 100Xp and captured either by a user supplied exhaust or an ATI Local Exhaust Module, if installed.
Aerosol Detection - Photometry:
The forward light scattering photometer is designed based on MIE scattering theory for analyzing particles dispersed in a fluid. This design includes a cylindrical housing enclosing a pair of axially spaced, spherical-surfaced, bi-convex lenses which provide axial focusing of the lamp image in an intermediate light scattering chamber through which the aerosol sample being analyzed is passed.
Comments
0 comments
Please sign in to leave a comment.