PARAMETER SETUP
To enter setup mode press <ENTER> then the UP arrow
L0 – Run
L1 – Audible Alarm
L2 – Display Rate
L3 – 100% Sample Time
L4 – Decimal Places
L5 – Internal Reference
L6 – Hour Meter
L7 – Factory Defaults
L8 – Software Revision
L9 – Bar graph
L10 – Display Intensity
Use the UP or DOWN arrow to select and <ENTER> to view setting
ALARM (ON)
- <ENTER>
- <ALARM> Led Flashes – (displayed indicating alarm is disabled)
- <ALARM> Last alarm set point displayed, use the UP or DOWN arrows to change
- <ENTER> Alarm set point entered and LED will stay lit
ALARM (OFF)
- <ENTER>
- <ALARM> Set point displayed
- <ALARM> displayed indicating alarm will be disabled
- <ENTER> Alarm disabled
INTERNAL REFERENCE
- 3-Way Valve to Clear
- <ENTER>
- <REF> P1 100.0 or last reference setting displayed, use the UP or DOWN arrows to change
- <ENTER> Scans 15 seconds, then 0 LED flashes
- <ENTER> Sets 0% – approx. 5 seconds Ready for Testing
SETTING 100%
- Sample Line Connected & 3-Way Valve to Upstream
- <ENTER>
- <100> Led Flashes
- <ENTER> Bar graph Scans, then 100 flashes and averages 100% for duration of L3 Setting, Zero LED flashes, Turn 3 – Way Valve to Clear
- <ENTER> Sets 0% – 5 seconds
- Ready for Testing
UPSTREAM CHALLENGE QUICK CHECK FEATURE – (Will function only if INT REF was established previously)
- 3-Way Valve to Upstream with Sample Line Connected (Now displays 100%)
- <REF> Bar graph scans 15 seconds
- Displays actual Upstream concentration for 5 seconds
- Bar graph scans and unit returns to 100% baseline